Surface Error of Concave Optical Surfaces
Optical Setups
Overview
Verifying the surface quality of concave optical surfaces such as lenses, mirrors, off-axis parabolas, and telescope optics is typically accomplished with either a Fizeau Laser Interferometer or a Twyman-Green Interferometer.
RECOMMENDED METHOD
Measure Concave Surface Error with a Fizeau Interferometer
Standard setup for measuring the surface quality of a concave optical surface.
Required Equipment
Typical setup for measuring concave surface error.
Required Components | |
AccuFiz Interferometer | More Information |
Transmission Sphere | Specifications |
5-Axis Base | Specifications |
Bayonet or 3-Jaw Mount | Specifications |
Other Considerations
Vertical Measurement
A vertical configuration can speed measurements, particularly for repeated measurements of the same type of component. The lightweight AccuFiz interferometer and lookdown stand make vertical measurement easy.
Measuring High Reflectors
For highly reflective optics the test beam returned to the system may be so bright that it washes out the interference fringes. A beam attenuator such as a pellicle can be inserted into the test path to improve beam contrast. A reference optic with attenuation coating can also be used.
A pellicle improves fringe contrast with a highly reflective test optic.
Measuring in Vibration
With a traditional, phase-shifting interferometer, measurement data is acquired sequentially—a long enough process that vibration and air turbulence will degrade or even prevent measurement. The interferometer and all mounts must be situated on an air table to isolate the system from vibration. Air isolation can be expensive—prohibitively expensive for very large optics or long measurement paths.
4D AccuFiz interferometers feature optional Dynamic Interferometry®, which enables all measurement data to be acquired simultaneously. Vibration is no longer an issue, and an isolation table is not needed.
Transmission Sphere Selection Guide
Learn how to choose the correct transmission sphere for your application.
ALTERNATE METHOD
Measure Concave Surface Error with a Short Coherence Fizeau Interferometer
The AccuFiz D short coherence Fizeau interferometer features an entirely on-axis design for excellent accuracy, particularly on highly curved surfaces. Dynamic Interferometry®, standard on the AccuFiz D, enables measurement without vibration isolation over cavity lengths up to 2 meters.
AccuFiz D short coherence Fizeau laser interferometer.
ALTERNATE METHOD
Measure Concave Surface Error with a Twyman-Green Interferometer
Standard setup for measuring the surface quality of a concave optical surface with a Twyman-Green interferometer.
A Twyman-Green laser interferometer is also well suited for verifying the surface quality of concave optical surfaces. It is the preferred method for verifying the surface quality of large optics.
Dynamic Interferometry, standard on PhaseCam Twyman-Green interferometers, enables long path measurements without vibration isolation—a substantial cost savings for meter-class optics.
Need Assistance with Measuring Concave Optics?
Contact our applications staff today for expert opinions on optical measurement and on measurement equipment.
Get In Touch
(520) 294-5600
Location
3280 E Hemisphere Loop, Ste 146
Tucson, AZ 85706
4Dinfo@ontoinnovation.com
Office Hours (Arizona Time)
Mon: 8am - 5pm
Tue: 8am - 5pm
Wed: 8am - 5pm
Thur: 8am - 5pm
Fri: 8am - 5pm
Sat: Closed
Sun: Closed